Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 32

Advertising
background image

ICS-2100 Ion Chromatography System

20

Doc. 065291-01 3/09

Suppressor

The suppressor reduces the eluent conductivity and enhances the
conductivity of the sample ions, thereby increasing detection sensitivity.

Either an AES

®

Atlas Electrolytic Suppressor, SRS

®

Self-Regenerating

Suppressor, or MMS

MicroMembrane

Suppressor can be used with

the ICS-2100. See

Section 2.4.9

for details about the suppressor.

Separator and Guard Columns

Both the separator and guard columns are packed with resin and perform
the separation of the sample ions. The main function of the guard column
is to trap contaminants and remove particulates that might damage the
separator column.

Column Heater

The column heater controls the temperature of the separator and guard
columns. The temperature can be set to between 30 and 60 °C; however,
it must be set to at least 5 °C above the ambient temperature. See

Section 2.4.8

for details about the column heater.

Eluent Valve

The eluent valve controls the flow from the eluent reservoir. The eluent
valve opens automatically when the pump is started and closes when the
pump is turned off.

Screen Adjustment Knob

Use this knob to change the LCD touch screen contrast.

Tubing Chases

The upper tubing chase routes tubing from the top cover to the component
panel. The lower tubing chase routes tubing from the component panel,
through the interior of the ICS-2100, to the rear panel.