136

Doc. 065291-01 3/09

6. Pull the valve straight out from the

component panel and begin pulling
the attached cable out of the opening
in the panel.

7. The cable is connected to a

matching cable inside the ICS-2100.
Continue pulling until the
connectors for the two cables are
outside the panel (see

Figure 5-27

).

Disconnect the cables.

Figure 5-27. Eluent Valve Cable

8. Remove the mounting plate on the

old valve by unscrewing and
removing the two screws. Save the
screws.

9. Thread the cable from the new

eluent valve through the mounting
plate.

10. Attach the mounting plate to the

new valve using the screws
removed in

Step 8

.

Figure 5-28. Removing the Eluent Valve Mounting Plate

11. Connect the cable from the new valve (P/N 057945) to the ICS-2100 cable.

Feed the cables back inside the ICS-2100.

Eluent Valve

Cable

Disconnect
cables

Eluent Valve

Remove
screws