5 • Service

Doc. 065291-01 3/09

159

Plumbing the EluGen Cartridge for Conditioning with Carbonate
Eluent

Figure 5-55

illustrates the liquid flow path during the EluGen cartridge

conditioning procedure with carbonate eluent.

Figure 5-55. EluGen Cartridge Plumbing for Conditioning with Carbonate Eluent

EGC

DEGAS

EGC

IN

Connected to

Gas Separator

Assembly

TO DEGAS

ELUENT OUT

TO SRS/AES

REGEN OUT

WASTE,

GAS

SEPARATOR

Connected to

Suppressor

Regen Out

Pre-plumbed

Pump Pulse

Damper

to

Inlet

Outlet

EGC VENT

Send to waste

EGC
OUT

EGC II

K CO

2

3

TO DEGAS

ELUENT IN

Back-

pressure

Coil

TO INJ

VALVE IN - P/

ELUENT OUT

To waste

14 MPa (2000 psi)
at 1.0 mL/min

Connected

Inject Valve

Port P (2)

to

EGC CO

Mixer

3

TO EGC CO3

MIXER IN

TO EGC CO3

MIXER OUT

TO INJ VALVE

PORT P

Temporarily

disconnect from

TO INJ VALVE IN - P/

ELUENT OUT

TO PUMP/

DAMPER