Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 198

Advertising
background image

ICS-2100 Ion Chromatography System

186

Doc. 065291-01 3/09

4. Remove the

EGC

VENT

line (see

Figure 5-73

) from

the EluGen cartridge
by unscrewing the
luer lock from the
luer adapter.

If necessary, use a
wrench to hold the
luer adapter in place
while unscrewing
the vent line.

Figure 5-73. EluGen Cartridge Vent Line and Luer Fitting

5. Install the luer cap (P/N 053981), provided with the EluGen cartridge,

onto the luer adapter. This prevents leaks from the vent opening when
you turn over the EluGen cartridge (see

Step 8

).

6. If an EPM is installed, lift up the bracket with the EPM attached and

remove it from the EGC holder.

7. Lift the EluGen cartridge straight up and out of the EGC holder.

8. Turn the cartridge upside down and set it in the EluGen cartridge

service area on top of the ICS-2100 (see

Figure 5-36

).

Figure 5-74. EluGen Cartridge in Service Area

Luer Lock

Luer Adapter

EGC Vent
Line