Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 91

Advertising
background image

4 • Troubleshooting

Doc. 065291-01 3/09

79

your application. Reset the eluent concentration and/or the flow rate to the
recommended values.

This error occurs if the flow rate is set to a value not supported by the eluent
generator. The ICS-2100 allows a flow rate range of 0.00 to 5.00 mL/min
without an eluent generator. With an eluent generator, the allowed range is
0.10 to 3.00 mL/min.The recommended operating range is 0.40 to
2.00 mL/min.

To troubleshoot:

Set the flow rate to a value within the allowed range.

This error occurs when the current applied to the EluGen cartridge exceeds
the maximum current allowed. Under these conditions, the EGC current is
turned off to prevent damage to the EGC. This error may occur if the liquid
flow to the cartridge is interrupted.

To troubleshoot:

See

Section 4.6

for troubleshooting when there is no flow from the pump.

This error occurs when the current applied to the EluGen cartridge exceeds
the maximum current allowed. The EGC current is turned off to prevent
damage to the EGC. This error may occur if the liquid flow to the cartridge is
interrupted.

To troubleshoot:

See

Section 4.6

for troubleshooting when there is no flow from the pump.

EGC-1 INVALID FLOW RATE ERROR

EGC-2 INVALID FLOW RATE ERROR

EGC-1 OVER CURRENT

EGC-2 OVER CURRENT

EGC-1 OVER VOLTAGE

EGC-2 OVER VOLTAGE