5 • Service

Doc. 065291-01 3/09

147

14. Locate the yellow

0.5 mL/min, 7 MPa
(1000 psi)
backpressure coil
(P/N 053765) in the
ICS-2100 Ship Kit
(P/N 064375).

15. Connect one end of

this coil to the
EluGen cartridge

OUTLET

port (see

Figure 5-41

). Leave

the other end of the
coil unconnected.

Figure 5-41. EluGen Cartridge Outlet Connection for Conditioning the Cartridge

NOTE The backpressure coil connection is temporary; it is

used only during the EluGen cartridge conditioning
procedure described in

“Conditioning the EluGen

Cartridge” on page 150

.

16. Turn the EluGen cartridge over, with the fittings facing down. Shake

the EluGen cartridge vigorously and tap it with the palm of your
hand 10 to 15 times. Check that all bubbles trapped in the eluent
generation chamber are dislodged.

The EluGen cartridge contains one of the following: a corrosive base
(KOH, LiOH, or NaOH), or a corrosive acid (MSA). Wear protective
eyewear and gloves when handling the cartridge.

La cartouche d'EluGen contient un de ce qui suit: une base corrosive
(KOH, LiOH, ou NaOH), ou un acide corrosif (MSA). Porter des
lunettes et des gants protectives en manipulant la cartouche.

Die EluGen-Kartusche enthält eine korrosive Base (KOH, LiOH oder
NaOH), oder eine korrosive Säure (MSA). Tragen Sie daher beim
Umgang mit der Kartusche eine Schutzbrille und Handschuhe.

Backpressure

Coil

OUTLET

Port