2 • Features

Doc. 065291-01 3/09

33

Select the Options tab (see

Figure 2-14

).

.

Figure 2-14. Server Configuration Properties: Degas Options

Degas Options

Always Off

: The degas pump is always off.

Always On

: The degas pump is always on. This setting is for test

purposes by a Dionex Service Representative. Do not use this setting
for routine operation.

Cycle

: The degas pump cycles on and off according to the times

specified in the

Cycle On

and

Off

fields.

Cycle On

specifies for how

long the degas pump runs during a cycle.

Cycle Off

specifies the time

between cycles.

Monitor

: (default mode) The ICS-2100 monitors the degas pressure

reading and turns the degas pump on and off as required.