2 • Features

Doc. 065291-01 3/09

41

EPM Electrolytic pH Modifier and EGC CO

3

Mixer

The EGC II K

2

CO

3

EluGen

Cartridge can be used with an
EPM Electrolytic pH Modifier
(P/N 063175) (see

Figure 2-17

)

and an EGC CO

3

Mixer

(P/N 061686) to produce a
carbonate/ bicarbonate mixture.
The carbonate/bicarbonate
mixture is used with carbonate-
based IonPac columns for
anion exchange separations.

Figure 2-17. ICS-2100 with an EluGen Cartridge and EPM Installed

After the cartridge generates carbonate eluent, the pH Modifier adjusts
the eluent concentration to produce the carbonate/bicarbonate mixture.
The EGC CO

3

Mixer provides enough mixing to produce a homogenous

solution of electrolytically-generated K

2

CO

3

and KHCO

3

eluent.

For more information about these products, refer to the EluGen cartridge
manual. Cartridge manuals are included on the Dionex Reference Library
CD-ROM (P/N 053891).

Generating Carbonate/Bicarbonate Eluent

If a K

2

CO

3

EluGen cartridge and an EPM Electrolytic pH Modifier are

installed (connected to EGC-1 and EGC-2, respectively), set EGC-1 to
the concentration of K

2

CO

3

required for your application and set EGC-2

to the concentration of KHCO

3

required. In order to achieve the desired

K

2

CO

3

/KHCO

3

eluent mixture, the K

2

CO

3

cartridge must generate the

total of the two setpoint concentrations.

For example, for a 3.50 mM K

2

CO

3

/1.00 mM KHCO

3

eluent, set EGC-1

to 3.50 mM and EGC-2 to 1.00 mM. When the eluent is generated, the
K

2

CO

3

cartridge generates 4.50 mM K

2

CO

3

, which is then modified by

EPM

EluGen
Cartridge