Flow Accuracy

<0.1% typical

Pressure Ripple

<1% at 13.8 MPa (2000 psi) and 1.0 mL/min

Eluent On/Off

Valve

Standard feature

Piston Seal

Wash

Dual-pump head; wash can be continuous when connected to rinse
solution supply

Pressure Alarm

Limits

Upper limit: 0 to 35 MPa or 0 to 5000 psi in one unit (MPa or
psi) increments

Lower limit: Up to one unit lower than upper limit

Vacuum Degas

Optional feature; automatic control

Eluent Bottles

Polypropylene

Eluent Bottle

Pressure

Not required

Leak Detection

Built-in optical sensor