C • ttl and relay control – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 259

Advertising
background image

C • TTL and Relay Control

Doc. 065291-01 3/09

247

Figure C-6. Example Connections for Stand-Alone Operation

ICS-2100 TTL Inputs

TTL IN 1 = Inject
TTL IN 2 = Autozero
TTL IN 3 = Pump On/Off
TTL IN 4 = Mark

AS TTL Inputs

TTL IN 1 = (unassigned)
TTL IN 2 = Start/Continue Schedule
TTL IN 3 = Tray/Temp. On/Off
TTL IN 4 = Therm. Compart. On/Off

System Function

Inject
Autozero

Start/Continue Schedule

Pump On/Off
Start Data Acquisition

AS TTL Out 1 to ICS-2100 TTL In 1

to AS TTL In 2

AS TTL Out 2 to ICS-2100 TTL In 2
AS RLY Out 2 to ICS-2100 TTL In 3
AS RLY Out 1 to Third-party DAC RLY In
Third-party ADC RLY OUT

Connection

Analog Data Transfer

AS Rear Panel Connections

ICS-2100 Rear Panel Connections

red: pos. 7,8,9

1

2

3

4

5

6

7

8

9 10 11 12

1

2

(+)

TTL OUT

1

2

TTL GND

(-)

1

2

3

4

TTL IN

(+)

1

2

RELAY

OUT

TTL 2 OUT (+)

TTL 1 OUT (+)

RLY 2 OUT

RLY 1 OUT

TTL 1 IN (+)

TTL 2 IN (+)

TTL 3 IN (+)

TTL 4 IN (+)

TTL
IN/OUT (-)

Third-Party Analog-to-Digital Converter and Controller

RELAY OUT

red pos.
1,3,5,6

red

black

red

black

Start/Continue Schedule

Inject

1

2

3

4

5

6

7

8

9

10

11

12

black pos.
4,11,12,12

Autozero

ANALOG

OUTPUT

+

-

ANALOG IN

Pump On/Off

black:
pos. 11,12,12

Start Acquisition

RELAY IN

Advertising