9. Enter the column heater calibration values in Chromeleon or on the ICS-2100

touch screen

DIAGNOSTIC

page as follows.

To enter the calibration values in Chromeleon:

a.

Open the Wellness Panel (see

Section 5.1.1

) and click Calibration

Details.

b. Under Column Heater Calibration Values, enter the Column Heater

Offset and Column Heater Slope values recorded in

Step 4

.

c.

In the field next to the Download button, verify that Current is selected
and then click Download to download the new calibration values to the
ICS-2100.

d. To ensure the new calibration values are recorded in the ICS-2100

memory, do not turn off the ICS-2100 power for at least 1 minute after
downloading the values.

To enter the calibration values on the ICS-2100 touch screen

DIAGNOSTIC

page:

a.

Go to the

DIAGNOSTIC

page and touch

Column Heater Cal Values

.

b. On the

COLUMN HEATER CAL VALUES

page (see

Figure 5-24

), enter the

Calibrated Offset

and

Calibrated Slope

values recorded in

Step 4

.