ICS-2100 Ion Chromatography System

78

Doc. 065291-01 3/09

This error occurs if the measured current is outside the expected range when
running the eluent generator calibration procedure.

To troubleshoot:

1. Review the procedure to verify that you set up the ICS-2100 correctly and

then rerun the calibration.

2. If the error message reappears, contact Dionex for assistance.

This error occurs when Chromeleon sends a command to set an eluent
generator parameter when the EGC is disconnected. The error also can occur
if you attempt to run the EGC verification test when the EGC is disconnected.

To troubleshoot:

1. Verify that the EluGen cartridge cable is connected to the

EGC-1

connector on the ICS-2100 top cover (see

Figure 2-7

).

2. If the error persists, contact Dionex for assistance. There may be a

malfunction in the EGC control electronics.

The maximum eluent concentration that can be set for a particular application
depends on the suppressor type, the EGC type, and the flow rate. This error
occurs when the selected concentration is too high for the current flow rate.

To troubleshoot:

Refer to the column manual, which is included on the Dionex Reference
Library CD-ROM (P/N 053891), for the concentration of eluent to use with

EGC-1 CALIBRATION ERROR

EGC-2 CALIBRATION ERROR

EGC-1 DISCONNECTED ERROR

EGC-2 DISCONNECTED ERROR

EGC-1 INVALID CONCENTRATION VS. FLOW RATE ERROR

EGC-2 INVALID CONCENTRATION VS. FLOW RATE ERROR