Hardware installation, Step 1 install 5200a, Location considerations – YSI 5200A User Manual

Page 14

Advertising
background image

YSI 5200A
26

YSI 5200A
27

Hardware Installation

Correctly install 5200A system components and peripheral devices to ensure accu-

rate data collection and reliable operation. First, determine what monitor, control,

and alarm functions the 5200A system will perform. Because system components

are wired to the 5200A instrument, it is important to determine where and how

components will be installed and configured. Preparing a system schematic (figure

3.1) is helpful.

Figure 3.1

Hardware installation steps (not all steps required for all applications):

Step 1 - install 5200A instrument

Step 2 - locate and install probe/cable assembly

Step 3 - ground 5200A

Step 4 - wire power

Step 5 - wire relays (output devices)

Step 6 - wire aux inputs

Step 7 - wire a network of 2 - 32 instruments

Step 8 - install and wire communication equipment

Step 9 - connect probe assembly

See Chapter 4 “Configuring 5200A Systems” for directions on programming 5200A

instruments.

Installation and Wiring

Tools and supplies needed for hardware installation:

Phillips-head screwdriver to remove 5200A front panel to access I/O board

and lower board assembly

necessary tools for cutting mounting holes in mounting brackets (rail

mount) and/or control panel (panel mount)

5/32” Allen wrench (panel mount only)

electric drill and bits for mounting 5200A (panel or rail) and for drilling

rubber grommets in bulkhead fittings

small flat-head screwdriver to terminate wires on the I/O board’s plug-

gable terminal strips

wrench to tighten the bulkhead fittings following system component

wiring

tweezers to replace gasket in YSI 5561 probe

wire strippers

distilled water for mixing DO membrane electrolyte solution

contactors may be required for relay wiring; see

contactor wiring - page

47

Optional accessories for hardware installation:

surface mount hardware (included)

rail mount hardware

panel mount hardware

weather shield hardware

serial to ethernet device

wiring for network wiring

wiring not supplied with components and/or peripheral devices

Step 1 Install 5200A

Select a mounting location and method to mount the 5200A.

Location Considerations

The 5200A must be located:

above any level where water damage can occur

away from extremely high or low temperature sources

away from vibrating surfaces

at least two (2) feet (.7 m) from any high voltage conduit

away from electromagnetic, radio, AC motor, transformer, or antennas

so the front panel can be fully opened and serviced

so system component wires are run as specified in this manual and ac-

cording to all local applicable electrical codes

Installation and Wiring