System menu – YSI 5200A User Manual

Page 42

Advertising
background image

YSI 5200A
82

YSI 5200A
83

After one minute, verify that the pH and temperature values show no

9.

significant change for approximately 30 seconds. Then, press

to

continue the calibration OR press

to escape calibration and return

to calibration menu without saving.

10.

Using the arrow keys to select numbers on the numeric keypad,

enter the pH buffer value. Press

to confirm each number selection.

Press

11.

to continue OR press

to escape the calibration and

return to calibration menu without saving.

For one-point calibration skip to step

12.

18.

Rinse the calibration cup and sensors, then fill the calibration cup with

13.

the 2nd buffer solution for a 2-point calibration.

Replace probe in calibration cup and let settle for at least 30 seconds.

14.

After one minute, verify the pH and temperature values show no significant

15.

change for approximately 30 seconds. Then, press

to continue the

calibration OR press

to escape calibration and return to calibration

menu without saving any of the calibration.

16.

Enter the value of the 2nd buffer solution. Press

to confirm

each number selection.
Press

17.

to continue OR press

to escape the calibration and

return to calibration menu without saving.
Press

18.

to save calibration and return to Run Screen.

Remove calibration storage cup from probe assembly.

19.

Rinse the sensor in tap or purified water before continuing to calibrate

20.

other sensors or reinstalling the probe guard.

Configuring the 5200A

System Menu

Menus → System

The System Menu contains 17 submenus. Use System submenus to configure non-
sensor and non-timer systems. System submenus are shown below. Press

at

the Run Screen to enter the Main Menu. Use

to scroll and highlight a System

submenu. Press

8

to select and enter a highlighted menu. Configuration changes

are saved when

is pressed.

Notes:

See

-

Appendix 3 “Menu Maps” for system defaults and ranges (where ap-

plicable) - starting on page 194.

Menu configurations are saved to the instrument only after

-

is

pressed and not when the change is made in the menu. If

is not

pressed, the changes made to the configuration will not be saved. Excep-

tion - Date and time and clear data logs are updated immediately to the
system; however, YSI recommends pressing

in this instance as

well in order to save any other changes that may have been made to the

configuration.

Configuring the 5200A