Reset System

Menus → System → Service→ Reset To Factory→ Reset System

Specific systems or all systems listed in the Reset System menu can be reset to fac-

tory defaults.

Use

1.

and

to scroll and highlight submenus.

Press

2.

to select.

When prompted, “Are you sure?”, select one of the following to confirm,

3.

Press

4.

to save configuration and return to Run Screen.

Reset Sensors

Menus → System → Service→ Reset To Factory→ Reset Sensors

Specific sensors, aux system, or all sensor and aux systems listed in the Reset Sensors

menu can be reset to factory default.

Use

1.

and

to scroll and highlight submenus.

Press

2.

to select.

When prompted, “Are you sure?”, select one of the following to confirm,

3.

Configuring the 5200A

Press

4.

to save configuration and return to Run Screen.

Note:

Perform a sensor calibration after resetting a sensor to factory default.

-

Reset Timers

Menus → System → Service→ Reset To Factory→ Reset Timers

A specific timer, feed timer, or all timers listed in the Reset Timers menu can be

reset to factory defaults.

Use

1.

and

to scroll and highlight submenus.

Press

2.

to select.

When prompted, “Are you sure?”, select one of the following to confirm,

3.

Press

4.

to save configuration and return to Run Screen.

Relay Test

Menus → System → Service→ Relay Test

Use the Relay Test menu to verify peripheral devices are wired correctly. Entering

this menu turns off any active relays and disables any sensor, aux, and timer system

relays from energizing regardless of value or condition. Systems are not reset when

the Relay Test menu is exited. To reset systems after exiting the Relay Test menu,

power cycle the 5200A. Example - an energized conductivity low control relay

(normal mode) is de-energized when entering the Relay Test menu. If the conduc-

tivity value is in control low range when the Relay Test menu is exited, the relay

will not re-energize until the system has a change of condition. The conductivity

system must reach set point, then fall in low control range again before energizing

the control low relay. It is strongly recommended to turn the power off, then on, at

the 5200A after exiting the Relay Test menu.