Sensor replacement, Charging dc batteries, Aquamanager database maintenance – YSI 5200A User Manual

Page 93: Replacement parts & accessories

Advertising
background image

YSI 5200A
184

YSI 5200A
185

Follow the individual sensor storage instructions listed in the following

3.

section.

4.

Sensor Storage

Temperature/Conductivity Sensor

No special precautions are required. The sensor can be stored dry or wet as long

as solutions in contact with the thermistor and conductivity electrodes are not cor-

rosive (for example, chlorine bleach). However, it is recommended that the sensor

be cleaned with the provided brush prior to long term storage.

pH and Combination pH/ORP Sensor

The key to sensor storage is to make certain that the reference electrode junction

does not dry out. Junctions which have been allowed to dry out due to improper

storage procedures can usually be rehydrated by soaking the sensor for several hours

(overnight is recommended) in a solution of 2 molar potassium chloride or pH

buffer 4. However, in some cases, the sensor may have been permanently damaged

by dehydration and will require replacement.

CAUTION: Do not store the sensor in distilled or deionized water as the sensor

may be damaged by exposure to this medium.

Remove the pH or pH/ORP sensor from the probe module.

1.

Seal the empty port with the provided port plug.

2.

Place the sensor in the storage vessel (plastic boot or bottle) that was on

3.

the sensor during shipment. The vessel should contain a solution which

is 2 molar potassium chloride or pH buffer 4.

Note - Make certain that the vessel is sealed to prevent evaporation of the storage

solution.

DO Sensor

The DO sensor should be stored dry with a dry membrane cap installed to protect

it.

Remove the used membrane cap.

1.

Rinse the DO sensor with distilled or deionized water.

2.

Dry the DO sensor and install a new, dry membrane cap.

3.

Sensor Replacement

All 5200A sensors are user replaceable. Refer to the instruction sheet provided with

the replacement sensor for detailed instructions.

Charging DC batteries

The 5200A does not charge batteries. Quality assurance maintenance procedures

should be established if batteries are to be fully powered when used as back-up

power source.

AquaManager Database Maintenance

See AquaManager Database Maintenance - page 171 and AquaManager Online Help

for additional information.

Replacement Parts & Accessories

See www.YSI.com for replacement part and accessory information and Appendix

5 of this manual - page 220.

Maintenance

Maintenance

Advertising