30 Second Aux and Sensor System Hold Off

Exiting the Aux and Sensor menus after making certain configuration changes places

the system in a 30 second hold. During the 30 second sensor hold, control and

alarm systems are not active. System(s) becomes active if, after 30 seconds, values

are reported in control or alarm range. This 30 second hold off allows sensors time

to stabilize before control and/or alarm systems become active.

Example - conductivity sensor menu is entered when dose up system is active and

control relay is energized. Dose up wait time is changed from 15 seconds to 45
seconds. When

to Run Screen is pressed, the conductivity system dose up

wait time is saved. Dose up relay de-energizes regardless of sensor value. Dose up

control relay re-energizes if, after 30 seconds, the value is still in control low range.

Dose up relay will energize for the new configured dose up time.

Reconfiguration of the following does not place the system in 30 second hold off:

All systems - enable/disable alarms;

-

DO system - display format, probe type, membrane type, manual ppt

-

input, elevation, temp source, temp display format, and low alarm set

point;

Conductivity system - Conductivity system;

-

Temperature system - display format, chiller on time, chiller on time units,

-

chiller hold off time, and chiller hold off time units.

Notes:

Configuration changes place only the system that has been reconfigured on

-

30 second hold. In the above conductivity example, only the conductivity

system has a 30 second hold off. Aux, pH, ORP, DO, and temperature

control systems are not affected. When configuration changes are made

to any system, all active alarm systems are suspended until returning to

Run Screen.

No “off” event log is created when energized system relays de-energize

-

due to 30 second sensor hold off.

Configuring the 5200A

Control icons and control messages continue to display during the 30

-

second sensor system hold at the Run Screen regardless if relays are

energized.

Enabling certain control/alarm functions when a system value is in con-

-

trol/alarm range requires that the 5200A be power cycled in order to reset

the system to properly control/alarm for the current condition.

Aux Setup

Menus →Sensor Setup → Aux 1-2

Aux systems control/alarm on (digital), control/alarm high, control/alarm low

(analog), hold off, and relay setting are configured in the Sensor Setup menu. Before

configuring Aux Sensor Setup menus, configure the Aux input type at the system

menu. See Configure Aux - page 98 Menus → System → Aux Configuration 1-2

menu. Aux Sensor Setup menu selections are based on the Aux input type config-

ured at system menu.

Notes:

For normal operation, the 5200A must be at the Run Screen. Alarm

-

functionality (alarm emails, activation of assigned alarm relays,

alarm icons, and buzzer) is suspended when 5200A is not at the Run

Screen.

There is a 30 second delay at power up before control, timer, and alarm

-

systems become active.

There is a 30 second sensor system hold off when sensor configuration is

-

changed. See 30 Second Sensor System Hold Off - page 132.

See

-

Menu Maps - page 194 for system defaults and ranges.

Aux Relay

Aux systems work differently than other sensors in regard to control and alarm

functionality because only one relay can be assigned to Aux systems. In addition,

there is no set point for analog aux systems. Four onboard relays can be assigned

for either aux control or alarm devices.

Notes:

The

-

menu will display energized aux relay only when alarms

are enabled.

An alarm icon is displayed at the Run Screen only when alarms are en-

-

abled. If alarms are disabled and an aux relay is energized, there is no

icon at the aux system on the Run Screen.

One relay can be assigned to multiple 5400 systems. It is essential that

-

the relay configuration be verified in order to avoid conflict of operation

Configuring the 5200A