Firmware can only be uploaded to an instrument using a local RS232 con-

-

nection. See Downloader - page 109.

On a network, install and configure RS232 or the serial to ethernet device

-

(hardware and wiring) on the master node only.

Email Alarming

Email alarming can be configured in one of two ways:

- from a PC running the AutoPoll Program that is included with AquaMan-

ager software (PC must have internet connection). This is referred to as

AquaManager email.

via a serial to ethernet* device with TCP/IP connection at master 5200A.

-

This is referred to as 5200A email.

*Serial to ethernet connection requires access to a SMTP server accessible by the

local network.

RS232 Communication

Use Category 5e patch cable from RS232 port at location H - figure 3.29 of I/O

board to PC using RJ45 to DB9 adapter.

Directions to Wire Direct RS232 PC Connection

WARNING: Disconnect external power to the unit before wiring

.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer

un câblage quelconque.

Perform steps 1 - 4 of 5200A installation.

1.

Open front panel -

2.

page 39.

Feed direct connect cable through drilled hole of rubber grommet and

3.

to location H.

Run high and low voltage cables through separate bulkhead and con-

4.

duit.

Connect one end of communication cable to RJ45 at location

5.

H. Connect

other end to PC com port using RJ45 to DB9 adapter.

Installation and Wiring

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5200A is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.29

Complete other wiring to 5200A.

6.

Close front panel -

7.

page 40.

Test RS232 communication.

8.

Notes:

- When using a USB-Serial Adapter, verify that the port number assigned

during driver installation is the same port configured when using Aqua-

Manager and/or the Firmware Updater.

When making RS232 cable runs greater than 100 feet (30.5 m), use an RS232

-

Line Extender/Booster. Mount the booster in an area where it is protected

from moisture.

Proper PC com port number must be configured when using AquaManager

-

or Firmware Updater programs.

Test RS232 Communications

Apply power to 5200A. Insert node into AquaManager program. Poll node. If

wired correctly, AquaManager will get 5200A configuration and logs.

Ethernet (TCP/IP) Module Communication

The optional serial to ethernet device must be installed on the 5200A control board

when using TCP/IP for communication. Complete installation instructions are

provided with the optional serial to ethernet device.

Installation and Wiring