Step 4 wire power, Wire ac power – YSI 5200A User Manual

Page 22

Advertising
background image

YSI 5200A
42

YSI 5200A
43

Perform steps 1 - 2 of 5200A installation.

1.

Open front panel -

2.

page 39.

Remove screw from location

3.

E.

Feed ground wire through drilled hole in rubber grommet and to location

4.

E on metal plate covering I/O Board. Ground wire gauge must conform

to all locally applicable electrical codes.

CAUTION

: Run high and low voltage cables through separate bulkhead and

conduit.

Terminate ground wire by returning screw at

5.

E over terminal connection

of ground wire.

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5200A is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.16

Terminate other end of the earth ground wire to a cold water pipe, steel

6.

rod, copper pipe, or earth ground.

Complete other wiring to 5200A.

7.

Close front panel -

8.

page 40.

Step 4 Wire Power

The 5200A is available as an AC or DC instrument. The DC version can only be

powered by DC power. The AC version can be powered by either AC or DC power.

If the AC version is properly wired and configured for both AC and DC power and

AC power fails, the power switches to 12VDC. If DC voltage drops below a user-

defined safe operating range, an alarm occurs.

Installation and Wiring

CAUTION: At 120VAC, the 5200A draws 25 watts (~0.2 amps). Use wiring

suitable for that load following all safety information and local electrical codes.

Back up power is recommended when operating with AC power. Backup power

can be supplied by 12VDC or UPS battery source.

Wire AC Power

The AC power option includes an integrated universal (worldwide) AC switching

power supply, allowing AC power from any AC receptacle 100-240VAC. AC power

wires are terminated at location L on I/O Board - figure 3.17.

Directions to Wire AC Power

WARNING: Disconnect external power to the unit before wiring.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant

d’effectuer un câblage quelconque.

Perform steps 1 - 3 of 5200A installation.

1.

Open front panel -

2.

page 39

Feed AC power cable through drilled hole in rubber grommet to location

3.

L. The terminal strip is removable.

CAUTION:

Run high and low voltage cables through separate bulkhead and

conduit.

Terminate AC power wires to pins “G” (ground), “L1”, and/or “L2/N”

4.

according to applicable local electrical codes. Use minimum 18 gauge,

3 wire power cord.

CAUTION:

It is recommended to power only the 5200A instrument with the

power cord to prevent exceeding the current draw of the AC power supply.

Installation and Wiring

Advertising