Service hold version, Aquamanager and the system menu – YSI 5200A User Manual

Page 63

Advertising
background image

YSI 5200A
124

YSI 5200A
125

Press

1.

to toggle relay on.

Press

2.

to toggle relay off.

Press

3.

to save configuration and return to Run Screen.

Note:

Timer and control icon(s) remain on the Run Screen until the system is

-

reset. Example - the timer icon remains at the systemwide location until

the end of the timer cycle that was stopped when the relay test menu was

entered even though that relay is not energized. To reset systems after

exiting the Relay Test menu, power cycle the 5200A.

Service Hold

Menus → System → Service Hold

Pressing

at the Run Screen puts the 5200A into Service Hold. The Service

Hold time is user selectable (1-255 sec, min, or hours). Service Hold turns off any

active relays and disables any sensor, aux, and timer system relays from energizing

regardless of value or condition. 5200A alarm emails are not sent during service

hold. After the service hold time expires or is cancelled, sensor systems are reset

after 30 seconds. Timer systems are not reset.

Use

1.

and

to scroll and highlight submenus.

Configuring the 5200A

Press

2.

to select.

Enable and configure submenus.

3.

Press

4.

to save configuration and return to Run Screen.

Version

Menus → System →Version

The Version menu contains information about the 5200A including: display and

control board firmware code version, control board temperature, and AC and DC

voltage.

Notes:

Firmware version information is required when contacting YSI sup-

-

port.

The clock chip temperature range starts at 0

-

oC. Values < 0oC are not valid

numbers.

Firmware date and time compilation are displayed below the version

-

number.

The normal operating temperature range will be a few degrees above

-

ambient.

AC powered units will run at a higher temperature than DC powered

-

units.

- To view current AC and DC power values press

to return to the

Run Screen and then return to the menu. Power information does not

refresh when the Version menu is open.

AquaManager and the System Menu

AquaManager users can configure most 5200A system menus from a PC. After poll-

ing the 5200A, use “Systems Management” from the node window to view and upload

5200A configurations. System menus are configured at the Setup and Input tabs.

Configuring the 5200A