Date/time display settings – YSI 5200A User Manual

Page 54

Advertising
background image

YSI 5200A
106

YSI 5200A
107

Press

7.

to return to Run Screen or

to System menu.

Notes:

- Logs cannot be deleted from the 5200A instrument using AquaMan-

ager.

Calibration and Configuration

-

logs cannot be cleared from 5200A.

Date/Time

Menus → System → Date/Time

It is important to configure the correct date and time in the 5200A. Four 10-event

daily timers and one feed timer run based on the 5200A date and time. Logs record

events based on the 5200A clock. Date and time display formats are user defined.

Enter Date/Time menu

1.

Menus → System → Date/Time

Verify 5200A current date and time.

2.

If necessary, use

3.

and

to scroll and highlight submenus.

Press

4.

to select.

Configure submenus.

5.

Press

6.

to save setting and return to Run Screen or press

to System menu.

Notes:

Changes to Date and Time menus take place immediately and are saved

-

prior to

.

Configured date and time formats are carried throughout menus and

-

logs.

Any active (with energized relay) timer system, including 10-event daily

-

timers, feed timers, and/or dosing timers, are NOT recalculated when

the 5200A time and/or date is changed. Once a timer relay has been

energized, it runs for the duration and does not reset based on changes

Configuring the 5200A

to the 5200A date and time. Example - timer 1, event 1 is configured to

come on at 9:00am for 10 minutes. Timer relay energizes at 9:00am. At

9:05am the 5200A time is changed to 10:05am (for daylight savings ad-

justment). The timer relay will de-energize after five additional minutes.

The 5200A time will be 10:10am. The event log will reflect the configured

5200A time when the event occurred. In this example, 9:00am for the

start and 10:10am for the end.

The time must be manually changed for daylight savings time adjustments.

-

Daylight savings note: Sensor data log records are not overwritten when

the clock is changed unless the instrument is power cycled after changing

the data and time. Example - the data log interval is set to 15 minutes.

Current instrument time is 6:00 am. The clock is adjusted to 5:00 am for

daylight savings time. The 5200A sensor log contains four 15 minute in-

terval records from 5:00 am to 6:00 am from prior to time change. Sensor

data logs are not overwritten with new values for the repeat of 5:00 am to

6:00 am 15 minute intervals unless the instrument is power cycled.

Menu configurations are saved to the system by pressing

-

to return

to the Run Screen and not when the change is made in the menu. Excep-

tion - Date and time and clear data logs are updated immediately to the
system, not when

is pressed.

Display Settings

Menus → System → Display Settings

Sensor and Aux conditions are displayed at the Run Screen. Use the Display Set-

tings menu to configure how many systems display on one page at the Run Screen.

Up to six systems can be displayed on one Run Screen page. Multiple Run Screen

pages scroll at user configured speed (1-65535 seconds).