Main menu calibration menu, Getting ready to calibrate, Main menu – YSI 5200A User Manual

Page 37: Calibration menu

Advertising
background image

YSI 5200A
72

YSI 5200A
73

Main Menu

Press

at the Run Screen to enter the Main Menu.

All system configuration is done through this menu. The four Main Menu submenus

are Calibration, System, Sensor Setup, and Timers.

Calibration Menu

Sensors must be user calibrated to ensure accurate measurements. Calibrate the

sensors in the Calibration Menu. Entering the Calibration Menu de-energizes all

active water sensor control and alarm relays.

There is a 30 second hold off period for control and alarm relays after pressing

or

at the calibration menu. The hold off period allows time for

the sensors to stabilize before processing readings for controlling and alarming

purposes.

Notes:

Energized timed sensor control relays (for example, conductivity, pH, &

-

temperature high) begin a new dose cycle after the 30 second hold off

time. For example, say the pH dose up is set to five minutes but the pH

low range relay de-energizes after dosing for only two minutes because

Configuring the 5200A

the calibration menu is entered. If the pH system remains in low range

control after the calibration is saved and the 30 second hold off period

has passed, then the pH low range relay will begin a new five minute

dose cycle. The dose cycle does not dose for the remainder of the timed

dose cycle that was active when the calibration menu was first entered.

Instead, it begins a new dose cycle.

Active Aux and Timer relays do not de-energize at calibration menu.

-

The temperature reading is displayed at Conductivity, ORP, and pH

-

calibration menus.

A “

-

Check probe” error message during a calibration indicates that the

sensor is unable to calibrate within the 5200A’s specifications. This may

indicate that the sensor needs serviced, the sensor is faulty, or improper

probe assembly setup or connection. See Connect Probe Assembly - page

60.

- The temperature sensor does not require calibration nor is it available.

User-configured sensor names appear at the top of the calibration

-

menu.

Energized alarm relays for all systems de-energize when the front panel

-

key is pressed to enter the calibration menu. Alarm systems are not reset
until

is pressed to return the instrument to the run screen.

Getting Ready to Calibrate

Use the calibration/storage cup, figure 4.5, for all 5562 sensor calibrations. To avoid

cross-contamination of solutions, rinse the calibration/storage cup between calibra-

tions. Additionally, rinse each sensor with a small amount of calibration solution

that can be discarded before placing the sensor in the calibration cup. Sensors

must be immersed in solution when calibrating Conductivity, pH, and ORP. The

recommended amount of calibration solution is provided in the sensor specific

calibration instructions.