YSI 5200A User Manual

Page 26

Advertising
background image

YSI 5200A
50

YSI 5200A
51

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5200A is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.21

Complete other wiring to 5200A.

5.

Close front panel -

6.

page 40.

Test aux systems.

7.

Test Auxiliary Input Devices

Apply power to 5200A. See Aux Setup - beginning on page 133 for configuration

information. Verify display, control, and/or alarm functionality based on user-

defined configuration.

Installation and Wiring

Wire Aux Digital Input

Install aux digital device according to manufacturer’s instructions.

1.

Connect switch wires to Aux Input 1 or 2. Connect one wire at (G) at

2.

K. Connect

second wire to (1) or (2) at K.

Float
Switch

Auxiliary Input
Connector

CAUTION

:

Do not input external voltage.

Figure 3.22

Wire Aux Analog Input

Install analog instrument according to manufacturer’s instructions.

1.

Connect analog instrument wires to Aux Input 1 or 2. Connect negative wire to

2.

ground (G) at

K

.

Connect positive wire to (1) or (2) at

K.

Instrument with analog
voltage or 4-20 mA output

Auxiliary Input
Connector

+ -

CAUTION

:

Observe correct polarity on analog input.

Figure 3.23

Installation and Wiring