background image

YSI 5200A
62

4

Configuring

5200A Systems

-Front panel keys
-Run Screen
-Navigating menus
-Menu structure

-Configuring 5200A systems

the 5200A. It is very important to follow the steps below to prevent damage to the

electronic components and to extend the life of the 5200A system.

Place solid rubber grommets in bulkhead fittings that have no electrical

1.

connections.

If using optional conduit fittings (YSI p/n 065926UL) in place of bulkhead

2.

fittings, industrial encapsulant (YSI p/n 065921 conduit sealer) must be

used to prevent moisture from entering the 5200A enclosure. Apply the

sealant after all wiring and connections are complete. Failure to use in-

dustrial encapsulant may result in damage to the 5200A.

Desiccant pack must be replaced whenever the 5200A enclosure is opened

3.

(Desiccant pack, YSI p/n 006506). Desiccant absorbs moisture captured

within the enclosure.

Ensure that the rubber gasket is seated between front panel and enclosure

4.

whenever enclosure is closed.

Replace the front panel ensuring that the ribbon cable is not trapped in

5.

the gasket channel before inserting mounting screws.

Tighten mounting screws, making sure not to cross thread. The screws are

6.

stainless steel and the receiving threads are brass. Do not over-tighten!

WARNING: The 5200A utilizes sensitive solid-state devices that can be dam-

aged by static shock. Installers must observe accepted ESD (Electro-Static

Discharge) procedures while connecting cabling to the 5200A I/O plate or

any other internal component or damage may result.

AVERTISSEMENT : le modèle 5200A utilise des appareils transistorisés sensibles

pouvant être endommagés par des décharges d’électricité statique. Les installateurs

doivent observer les procédures de protection acceptées contre les décharges

électrostatiques lors du câblage à la carte d’E/S du modèle 5200 ou à tout autre

composant interne afin d’éviter tout risque de dommage.