Menu - page 124 for additional information. No “off” event is created

for system relays that de-energize when 5200A is placed in downloader.

Systems are reset when power is cycled after downloader.

All logs and instrument configuration are retained when 5200A firmware

-

is updated.

Display board code loads first, followed by control board code. The display

-

board screen flickers when display board code is loading. The display stops

flickering when the control board code is loading. The Firmware Updater

progress bar will complete two full scrolls during firmware updating.

Total download time is approximately nine minutes; eight minutes to

-

upload display board firmware and approximately one minute to upload

control board firmware.

- Autopoll must be shut down when firmware is uploaded to 5200A.

Firmware Updater Messages

The Firmware Updater program may respond with one of the following messages

to indicate that the upload did not occur.

“Timed out” or “Could not find Unit” messages indicate a communication problem

between PC and 5200A. Reconfirm Communication menu configuration, com

port settings, and RS232 connections. (A power cycle will be required to access

the Communications menu.) Place 5200A in downloader mode again and select

“yes” to retry at Firmware Updater error message window. If the upload is still

unsuccessful, it is possible that the 5200A is not in downloader mode. See Force

Downloader - page 113.

“Unit is Up-To-Date” indicates that the 5200A has most recent firmware and the

upload is unnecessary.

Configuring the 5200A

Force Firmware Download

A status LED on the I/O Board must be visible to determine if the 5200A is in

downloader mode. The status LED flashes about 10 times in 5 seconds when in

downloader mode. When not in downloader mode, the status LED flashes more

quickly, about 15 flashes in 5 seconds. The LED at the “status/reset” location of the

I/O Board indicates the current status of 5200A boards.

Directions to Force 5200A into Downloader Mode

1.

WARNING: Disconnect external power to 5200A. (AVERTISSEMENT :

déconnectez l’alimentation externe de l’unité.)

Open front panel -

2.

page 39 .

Depress reset button on I/O board (see figure below) and hold down the

3.

enter key on the front panel. Use a paper clip or other sturdy piece of wire

to access the reset button through the I/O metal plate. Ensure that you

are only touching the reset button on the I/O board and the enter key,

then turn power on. The Downloader (bootstrap loader) screen should

appear on the instrument.

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5200A is po

w

er

ed

. S

er

ious damage can oc

cur

.

Proceed with uploading firmware with the Firmware Updater utility.

4.

See

Downloader - page 109.

General Alarm

Menus → System → General Alarm →

The General Alarm triggers the buzzer, general alarm relay, general alarm icon, and

5200A emails. Sensor alarms, aux alarms, slave failure (networks), and low battery

alarms trigger the General Alarm when enabled. If individual system alarms are