Directions to Enable Salinity System

Connect probe assembly -

1.

page 60

Enter Salinity Sensor Setup menu

2.

Menus →Sensor Setup→salinity

Select

3.

or

to turn off or on.

AquaManager and the Sensor Setup Menu

All Sensor Setup menus can be configured using AquaManager. After polling the

5200A node, use “Systems Management” to view and upload 5200A configura-

tion.

Configuring the 5200A

Notes

- Always manually poll the 5200A after uploading configuration changes

and verify that the uploaded information is retained. When configuration

changes are uploaded using AquaManager at the same time configura-

tion changes are being made by a user in the menus on the 5200A,

AquaManager uploads may not be saved depending on the timing of

keystrokes at the 5200A.

- The Run Screen must be displayed during normal operation. If the

5200A is not at the run screen during AquaManager uploading, pressing

at the 5200A will only save certain configuration changes.

See Chapter 5 AquaManager - page 166 and AquaManager online help for additional

information.

Timers

The 5200A has four daily ten event timers. Timers can be configured to control

various hardware devices. In addition, there is one Feed Smart

tm

feed timer.

Notes:

Timer system

-

must be enabled in order to energize the configured

timer relay.

If a timer is active (relay energized), the timer is NOT recalculated when

-

the 5200A time and/or date is changed. Once a timer has been turned

on, it runs for it’s configured duration and does not reset it’s run time

based on changes made to 5200A time. Example - Timer 1, event 1 is

configured to come on at 9:00am for 10 minutes. Timer relay energizes

at 9:00am. At 9:05am, the 5200A unit time is changed to 10:05am (for

daylight savings adjustment). The timer relay will de-energize after five

additional minutes. Unit time will be 10:10am. The event log will reflect

the unit time when the event occurred. In this example, 9:00am for the

start time and 10:10am for the end time.

If a timer is active (relay energized) and the 5200A is put on service hold,

-

the relay will not reenergize after service hold expires regardless if timer

time has not expired. There is only one “on” time for all timers. Turn the

5200A off and then on to reset the timer system. Timer relay will energize

for timer system configured time duration and not for the remainder of

time prior to service hold. Example - timer relay 1 is configured to start

at 11:00 am for one hour. Service hold is enabled and configured for 5

minutes. Timer relay energizes at 11:00 am. At 11:15 am, the 5200A is

put into service hold. Timer relay 1 de-energizes. At 11:20 am, the service

hold time expires. The 5200A is power cycled and timer relay energizes

Configuring the 5200A