YSI 5200A User Manual

Page 96

Advertising
background image

YSI 5200A
190

YSI 5200A
191

Issue

possible cause

possible solution

no sensor value on

screen

sensor not enabled

See Sensor Setup - page 127.

display setting configuration

set for multiple Run Screens

See Display Setting -

page107.

“check probe” message at

sensor calibration

improper placement of probe

assembly, incorrect probe as-

sembly connection, improper

placement of 5200A, bad sen-

sor, and/or bad membrane

See Location Considerations

Probe Assembly - page 33.

See Connect Probe Assembly

- page 60.

See Location Consider-

ations 5200A - page 27.

clean sensor, calibrate

sensor, test sensor outside

system stream in solution

of known value, replace

membrane and/or replace

sensor if necessary

Analog aux value is

overlapping to adjoining

sensor window

1X1 & 2X2 display format

setting cannot display values

over 7 digits in aux sensor

window

Adjust aux display format

to lower resolution i.e. from

0.00 format to 0 format. See

also Notes: Display Screen

Limitations - page 108.

Relay (Outputs) Troubleshooting

Disable control, alarm, and timer relays when troubleshooting relay

-

outputs.

Verify 5200A wiring and configuration.

-

Issue

possible cause

possible solution

relay does not energize

incorrect sensor configura-

tion

See Set Points, Controls, and

Alarm - page 128.

relay not enabled

incorrect timer configuration See Timers - page 147.
30 second delay at power up

and/or sensor configuration

change

30 seconds delay - see

page128 and page132..

peripheral device is wired

incorrectly

See Wire Relays - page 46

and Test Relays - page48.

Troubleshooting

Sensor Troubleshooting continued

Issue

possible cause

possible solution

relays are not enabled for

control/alarm/timer system

Verify menu configuration

(relay enable) for control/

alarm/timer system.

See sensor system enable -

page128.

See timer system enable

-page147.

temp (high), pH, and or

conductivity relays do

not de-energize when set

point values are read

sensor is configured for

timed control

Refer to index Control

Value

Email Alarm Troubleshooting

Issue

possible cause

possible solution

no 5200A email alarm

message received

5200A display is not at Run

Screen - all alarming is sus-

pended when in menus

exit to run screen to reset

alarm email alarm system

Serial to ethernet device did

not initialize properly. Check

the event log for “unable to

initialize error”.

disconnect power to 5200A,

verify installation of serial

to ethernet device on lower

board and repower

faulty serial to ethernet

device

replace serial to ethernet

device

Are the alarms enabled in

sensor menu? Is the general

alarm enabled?

See Sensor Setup Alarms

Values - page 129.

see General Alarm - page

113.

no email alarm received

for slave alarm condition

Are slaves enabled on net-

work?

enable slaves at Menus →

System → Communications

→ Network →Master

no test email alarm mes-

sage received

exit to run screen from com-

munication menu not done

exit to run screen to process

test email request

Troubleshooting

Relay (outputs) Troubleshooting continued