Troubleshooting

Network Troubleshooting

Issue

possible cause

possible solution

no communication to

networked unit(s)

Are master/slave menus

properly configured?

See Networking - page 93.

Is termination relay en-

abled in 1st & last 5200A on

network?

See Termination Relay -

page 95.

Is proper communication

method configured and

wired properly?

See Networking - page 93.

See Wire Communication

Equipment - page 57.

Are slaves enabled and recog-

nized at master?

enable slaves at Menus →

System → Communications

→ Network →Master

more than one device is

configured as a master

verify configuration at

Menus → System → Com-

munications → Network

→Master/Slave menus on

all nodes.

slow network perfor-

mance

nicked wires;

improper wiring

Verify network wiring -

See Wire Network - page 52

5200A Systems Troubleshooting

Issue

possible cause

possible solution

time and date lost at

power up

bad lithium battery

replace lithium battery

no events in event log

event logging not enabled

See Event Logging - page

104.

no data in data log

data log interval time has not

passed

See Data Logging - page102.

cannot view specific data

log records at 5200A

Data log record has older

date and time-stamp than

the 25 most current data log

records available at 5200A

sensor log.

See Data Logging - page102.

Troubleshooting

Firmware Updater Troubleshooting

Issue

possible cause

possible solution

Device “timed out” error incorrect configuration in

5200A communication menu

See Device Timed Out er-

ror - page 112.

incorrect wiring of RS232

connection

See Device Timed Out er-

ror - page 112.

5200A does not go

into downloader mode

when “program device”

is selected at Firmware

Updater

Manually put 5200A in

downloader and retry firm-

ware updater upload.

Menus → System → Down-

loader
See Force Downloader -

page 113.

unit does not upload new

code at downloader

unit is not in downloader

mode

See Downloader - page 109..

RS232 connection is incor-

rect

See Wire Communications -

RS232 - page 58.

Feed Timer Troubleshooting

Issue

Understanding how the 5200A calculates sensor reduction

time

Sensor reduction % is

not accurate based on

displayed sensor value(s).

The 5200A displays temperature sensor values to the 10ths

and DO values to 100ths. In reality the 5200A temperature

and DO sensor system values are stored at a much greater

resolution (out to 64 bits). It is the stored values, not the dis-

played values, that are used to calculate the feed timer sensor

reduction.

AquaManager Troubleshooting

Refer to AquaManager online Help for all AquaManager operation and troubleshooting

information.