The Feed Timer can be set to operate for any or all of the days of the week and

includes continuous or timed feeding modes, FCR (feed conversion ratio), and

sensor options.

The following definitions are used in regard to the Feed Smart

tm

Timer system:

feed cycle - amount of time that all feeds occur in:

-

continuous - feedings occur 24 hours/day

-

timed - length of time between start and end time

-

feed duration - length of time feeder relay is energized;

-

feed interval - length of time between feedings.

-

Optional Sensor control reduces the amount of food dispensed if DO, pH, or

temperature measurements are “out of range” per user defined configuration. The

FCR (Feed Conversion Ratio) feature automatically computes and increases the

daily amount of food dispensed per FCR configuration.

Notes:

It is important that the 5200A be configured with valid data. The feeder

-

system will not operate properly if, for example, the daily weight of food

is greater than can be dispensed in the defined feed cycle.

Disable the feed timer system when changing configuration and/or when

-

performing a calibration. Enable the feed timer system after configura-

tion changes and/or a calibration has been done. The feed timer system

recalculates duration and interval times based on the new configuration.
Press

after disabling the feed timer system and before making

configuration/calibration changes.

Feed Mode

Continuous

Menus →Timers→Feed Smart Timer→Feeder→Feed Mode →Continuous

When the feed timer is configured for continuous feed mode, feedings occur through-

out the day. Feed duration and interval are determined by the daily total weight of

food, number of feedings, and test calibration data. Optional Sensor control and

FCR configuration can also effect feed duration.

Configuring the 5200A

Example of Feed Timer Configuration and Test Calibration Data:

Feed Mode
C
ontinuous (86500 seconds in 24 hours), 
feed timer sensor control is disabled

daily weight
500 units

daily feedings
100

test calibration 
data
10 units per 30 
seconds

The 5200A determines feed duration and feed interval for the above feed cycle as follows:
units fed per feeding:

feed time duration for each feeding:
determine total time feed relay will be 
energized over 24 hour period:

15 units X 100 feeds = 1500 seconds

determine feed interval:

500 

÷ 100 = 5 units per feed

if 10 units feed in 30 seconds

, then 5 units will feed in 

15 seconds
30x5 

ч 10 = 15

86500 ‐ 1500 = 84900  

ч = 100 (number of feeds) = 849 

seconds¸ 60 = 14.15 minutes, or 14 minutes and 9 

Notes:

Once Continuous feed has been configured, the 5200A determines any

-

remaining feeds for the 24 hour period ending at 23:59. In continuous

feed mode, the feed cycle always begins at 00:00 (midnight). Start and

End time fields are not available for Continuous feed mode.

In continuous feed mode, feeding(s) will end after the last completed feed-

-

ing that ends prior to 11:59 pm (23:59) of an enabled day when crossing

midnight to a disabled day. Feeding(s) will resume at 12:00 am (00:00)

for the next enabled day.

Timed

Menus →Timers→Feed Smart Timer→Feeder→Feed Mode →Timed

Feedings occur only during the start and end times on “enabled” days of the week.

In timed mode, the 1st feeding occurs at the start time of the feed cycle. The last

feeding completes by the end time. Feed duration and interval are determined by

the daily total weight of food, number of feedings, and test calibration data. Optional

Sensor control and FCR configuration also effect feed duration.