Timed and normal control, Directions to configure ph sensor – YSI 5200A User Manual

Page 72

Advertising
background image

YSI 5200A
142

YSI 5200A
143

Press

6.

to save configuration and return to Run Screen.

Notes:

To configure a negative set point, control, and alarm values enter “-” after,

-

not before, the numeric digit(s).

See 30 Second Sensor Hold Off

-

- page132.

Only whole numbers can configured for ORP set point, control, and

-

alarm values.

pH

Menus →Sensor Setup→pH

See Set Points, Control, and Alarms - page 128 before configuring the pH system.

Timed and Normal Control

pH high and low control can operate in two different modes, normal or timed. In the

normal mode, control relays are energized until the set point value is reached. The

normal mode can be used for either low control, high control, or both. For normal

control, configure dose up/dose down to 0. The dwell time will be ignored.

In the timed mode, an energized control relay is energized for the dose time, re-

gardless if the sensor value has returned to the set point value. After the dose time

expires, the control relay de-energizes for the configured dwell time. After the dwell

time expires and if the value is still in control range, the control relay will reenergize.

A dwell time must be configured in the timed mode or the control relay will never

de-energize regardless of sensor value.

Configuring the 5200A

Directions to Configure pH sensor

Connect probe assembly -

1.

page 60.

Enter pH Sensor Setup menu

2.

Menus →Sensor Setup→pH

Use

3.

and

to scroll and highlight submenus.

Press

4.

to select.

Configure submenus.

5.

Press

6.

to save configuration and return to Run Screen.

Notes:

pH set point, control, and alarm values can be up to five (5) characters,

-

including the decimal point. Values truncate to 100ths when no 10s

digit is configured.

See 30 Second Sensor Hold Off

-

- page132.

Active dosing timers are reset after systemwide events. Systemwide events

-

de-energize relays. Example - pH dose up relay is configured for two (2)

minutes. Energized dose up relay de-energizes when 5200A goes into Svr

Hold. The dose up relay had been energized for 15 seconds before the

Svr Hold occurs. Svr Hold time is one (1) minute. After Svr Hold time

expires, the control system is reset. If the sensor is reporting a control

low condition, the dose on time will be for two (2) minutes and not the

remaining 45 seconds of the dose up time prior to Svr Hold.

pH Control Low Range message remains at the

-

menu when

the pH system is disabled and when a control low relay is energized.

Power cycle the 5200A to clear the control low range message at the

Msg’s menu.

Configuring the 5200A

Advertising