Input System Specifications

Aux Systems

There are 2 Aux input systems that allow digital or analog (0-

1.0VDC, 0-5.0VDC, or 4-20mA) configuration.

Aux System

Type

Digital

Analog

Range

dry contact open/

close

0 -1.0 VDC, 0-5.0 VDC, or 4-20mA

Accuracy

n/a

0-1.0VDC: ± 3% ± 10 mV

0-5.0VDC: ± 3% ±15 mV

4-20 mA: ± 3% ±100μA

Resolution

n/a

user selectable (0), (0 .0), or (0.00)

Probe assembly 5562 (DO, temperature, pH, ORP, and conductivity sensors)

Probe assembly 5561 (DO and temperature sensors)
Cable grade

military-spec (MS) connector

Cable lengths

4, 10, 20 m (special order up to 100m for DO and temp only)

Temperature
Sensor type

10,000 ohm @ 25°C thermistor

Range

0-45°C (32-113°F)

Accuracy

± 0.2°C

Resolution

0.1°C

Dissolved Oxygen
Sensor type

Membrane covered, steady-state polarographic

Range

0 to 500% air saturation

0 to 60 mg/L

Accuracy

±2% or ± 0.2 mg/L, whichever is greater for 0-200%

±6% of reading for 200-500%

Resolution

0.1% (0.0-99.9%); 1% (> 99.9)

0.01 mg/L (0-10 mg/L); 0.1 mg/L (10-60 mg/L)

Response Time

95% of reading in 17 seconds for 2.0 mil PE membrane;

95% of reading in 36 seconds for 2.0 mil Teflon membrane.

Allow approximately 10-15 minutes after 5200A is powered

on, or the DO sensor is serviced, for the polarographic sensor

to stabilize.

Introduction

Flow requirements minimum flow rate of 3 inches per second or 7.62 cm/second

for 2.0 mil PE membrane; 6 inches per second or 15.24 cm/

second for 2.0 mil Teflon membrane

Other

Automatic salinity compensation for DO readings when a

conductivity sensor is in use and enabled. Manual salinity

compensation for DO readings when a conductivity sensor is

not in use.

pH
Range

0.00 to 14.00 pH

Accuracy

± 0.1 pH

Resolution

0.01 pH

ORP
Range

-2000 mV to +2000 mV

Accuracy

± 20 mV

Resolution

1 mV

Conductivity
Sensor type

4 electrode, Cell constant of 5.0 ±5%

Ranges*

0 to 200 μS

0 to 2000 μS

0 to 20 mS

0 to 200 mS

Accuracy

0 μS to 100 mS, ±0.5% of range

100 mS to 200 mS, ± 1.0% of reading

Resolution

range dependent

0.1 μS (0 to 200 μS)

1.0 μS (

0 to 2000 μS)

0.01 mS (0 to 20 mS)

0.1 mS (0 to 200 mS)

Salinity
Parameter Source

1) calculated from temperature & conductivity systems when

enabled

OR

2) user ppt input in DO sensor setup menu

Range

0-80 ppt

Note: When not using a conductivity sensor, the manual

input range is 0-50 ppt.