YSI 5200A User Manual

Page 49

Advertising
background image

YSI 5200A
96

YSI 5200A
97

Send Test Email Msg

Menus → System → Communications → Send Test Email Msg

Send a test email to verify proper configuration of ethernet menus. Test emails are

only transmitted from configured master 5200As.

Note:

Exit to Run Screen is required before the test email is sent.

-

Example of a 5200A test email sent from an ethernet device:

Subject Line: 02/04/10 03:50:05 PM YSI5200A Email Alarm Test Message

From: [email protected]

Sent: Wed, Feb 04, 2010 3:49 pm

To: [email protected]

CC: [email protected]

Email Retry Delay

Menus → System → Communications → Email Retry Delay

Configure the Email Retry Delay for the time interval that emails should be retrans-

mitted after the initial alarm email is sent. The initial alarm email is sent immediately

after the general alarm hold off time or after the aux hold off time. Emails will be

retransmitted until one of the bottom front panel softkeys is pressed, regardless if

the alarm condition still exists.

Note:

On a network, emails are retransmitted according to the slave retry delay

-

configuration of the slave in alarm and not the master’s email retry delay

configuration. Please note, a slave Email Retry Delay menu will display

(Not available) because of its slave configuration; however, this feature

is available and has a default value of 15 minutes that can be configured

by the user.

The Retry Delay should be configured at intervals ≥ 15 minutes.

-

AquaManager alarm emails are retransmitted based on the 5200A slave(s)

-

and master email retry delay configuration.

Baud Rate RS232

Menus → System → Communications → Baud Rate

Configure the Baud rate (for RS232 connections only) in the 5200A to match the

baud rate configured in AquaManager. Use the default rate of 115200 for fastest

communication speed.

Configuring the 5200A

Enter Network menu

1.

Menus → System → Communications → Network →Baud Rate

Press

2.

or

to scroll and highlight a selection.

Press

3.

to confirm selection.

Press

4.

to save configuration and return to Run Screen.

Force Serial Port

Menus → System → Communications → Force Serial Port

Enable the Force Serial Port when the 5200A is communicating via RS232 (direct)

to a PC and when the 5200A has a serial to ethernet device.

Wire RS232 connection -

1.

page 58

Enter Network menu

2.

Menus → System → Communications → Force Serial Port

Use the

3.

softkeys to configure off or on.

Press

4.

to save configuration and return to Run Screen.

Notes:

RS232 connection is required to upload firmware using the Firmware

-

Updater.

The RS232 icon

-

blinks at the Run Screen when in Force Serial Port

mode.

Configuring the 5200A