Validation error and autofix – YSI 5200A User Manual

Page 66

Advertising
background image

YSI 5200A
130

YSI 5200A
131

It is important to keep the 5200A at the Run Screen during normal

-

operation. Alarm functionality (alarm emails, activation of enabled

alarm relays, alarm icons, and buzzer) is suspended when 5200A is

not at Run Screen.

Alarm icon(s) display at the Run Screen when alarm condition(s) exists

-

for longer than the general alarm hold off time. See Icons - page 67.

Press

-

at Run Screen to display active sensor system relays.

Press

-

to display events created by control and alarm relays - see

Event Log - page 104.

Alarm relays are not energized if the 5200A sensor value reads “Over”

-

or “Under.”

See Wire Relays

-

- page 46 for information on wiring alarm output de-

vices.

Figure 4.7 shows how the 5200A operates based on a sensor setup configuration.

In this example, the DO range low 1 and range low 2 relays energize at ≤ 5.0 mg/l

and ≤4.0 mg/l respectively. DO range high 1 and range high 2 relays energize at ≥

8.5

mg/l and ≥ 8.7 mg/l respectively. The alarm system becomes active at ≤3.5 mg/l

(low alarm) or ≥ 8.9 mg/l (high alarm).

Range #1

Low

Alarm

ON

High

Alarm

ON

High

Range #1

ON

Optimal

S

etpoint

Value

Lower

Parameter

Values

High

Alarm

Setpoint

Low

Alarm

Setpoint

Higher

Parameter

Values

DO Range #2

High

Range #2

ON

Low

Range #1

ON

Low

Range #2

ON

Acceptable
Operating
Range

Conductivity/ORP/pH & Temperature have 1 Control Range

DO has 2 Control Ranges

3.5 mg/l 4.0 5.0 8.1 8.5

8.7 8.9

Figure 4.7

Validation Error and Autofix

When an invalid value is entered at the numeric keypad, a Validation Error window

displays. An invalid value is either a value outside the range of the sensor system

or a value that breaches other settings within the sensor system. Autofix occurs

when a high or low control value is entered that breaches the set point value. See

examples below.

Configuring the 5200A

Low range control value must be < set point value. High range control value must

be > set point value. Low alarm value must be < low control value. High alarm value

must be > high control value.

Example 1 - entering “63” for a DO set point results in a validation error because the

valid range for the DO sensor system is 0.00 to 60.0 mg/L. The user must configure

the value within the operating range. In addition, the value cannot breach respective

control and alarm system values.

Example 2 - configuring set point value of 8.00 mg/L with range low value of 8.00

mg/L results in validation error. The valid set point range is displayed in validation er-
ror message. Press

to return to numeric keypad and to enter a valid value.

Autofix occurs when a high or low control value is entered that breaches the set point

value. Autofix changes the high or low control value by one least significant digit

away from the set point value. The set point value is never adjusted.

Example 3 - Entering 8.00 mg/L as the range low value when the set point value is

8.00 mg/L results in an auto fix of the control value. The invalid control value is
displayed at the sensor setup menu. Pressing

is not an option after saving

a value at numeric keypad. Press

from the sensor setup menu to continue

with the autofix. A validation error window is displayed to show that the autofix has
occurred. Press

to return to sensor setup menu. Verify that the autofixed

values are configured for user’s specific application and reconfigure as necessary.
Press

to save configuration and return to Run Screen.

Configuring the 5200A

Advertising