YSI 5200A User Manual

Page 20

Advertising
background image

YSI 5200A
38

YSI 5200A
39

Figure 3.13

Accessing the I/O Board

The I/O board is accessible when the 5200A front panel is removed. The letters on

the metal plate attached to the I/O board identify pluggable terminal strips on the

I/O board. Letters on the metal plate correspond to the wiring diagram located

on the inside of the front panel - figure 3.14. A ribbon cable connects the display

board (mounted in the front panel) to the I/O board. When the 5200A front panel

is removed, a lanyard protects against stress on the ribbon cable. See Open Front

Panel to Access I/O Board - page 39.

Installation and Wiring

Figure 3.14

Open Front Panel to Access I/O Board

WARNING: Disconnect external power to the unit before doing any wir-

ing.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant

d’ouvrir le panneau avant.

Loosen the 4 mounting screws connecting the front panel to the back

1.

enclosure. Captive screws remain in front panel after loosening.

Carefully lift the 5200A front panel from the back enclosure. Front panel

2.

should be placed to the left of enclosure. Allow the lanyard to hold the

weight of the front panel. The rubber gasket seated in the front panel

should remain in the front panel lid.

Installation and Wiring

Advertising