Serial number service – YSI 5200A User Manual

Page 60

Advertising
background image

YSI 5200A
118

YSI 5200A
119

Disabling the Power Fail System will disable AC power events and disable the General

Alarm in the event of low battery for DC power. Disable Power Fail AC Monitor

when powering 5200A by DC power source.

Enable Low Battery Monitor to trigger the General Alarm in the event DC voltage is

≤ to low battery trigger. The Low Battery Alarm Hold Off time is five (5) seconds and

cannot be user configured. Once the low battery general alarm has been acknowl-

edged by pressing one of 4 softkeys, it is not reset. See General Alarm - page 113.

Verify AC and DC power supplies are properly wired and powered -

1.

page

42.

Enter Power Fail Alarm menu and submenus

2.

Menus → System → Power Fail

Use

3.

and

to scroll and highlight submenus.

Press

4.

to select.

Enable and configure submenus.

5.

Press

6.

to save configuration and return to Run Screen.

Notes:

Press

-

to view AC power fail events.

-

icon at the Run Screen indicates that the 5200A DC power is ≥ Low

Battery Trigger value with Low Battery Monitor is enabled.

Configuring the 5200A

-

icon at Run Screen indicates DC voltage ≤ the Low Battery Trigger

when Low Battery Monitor is enabled.

Low battery alarm is not reset when front panel softkey is pressed.

-

AC power fail does not generate a General Alarm.

-

Serial Number

Menus → System → Serial Number

The serial number is displayed at the serial number menu. The serial number is

required when contacting YSI customer service. The serial number is also neces-

sary to configure AquaManager. 5200A serial numbers begin with “53-”followed

by seven digits. Serial numbers are generated at the factory and cannot be changed

by the user.

Service

Menus → System → Service

The Service menu includes submenus to configure and Reset Clean Probe Timer,

Reset Systems to factory defaults, and test Relay(s).

Configuring the 5200A