Wire dc power and backup power – YSI 5200A User Manual

Page 23

Advertising
background image

YSI 5200A
44

YSI 5200A
45

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5200A is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.17

Complete other wiring to 5200A.

5.

Close front panel -

6.

page 40.

Test AC power.

7.

Test AC Power

Turn on main power. The enabled sensor values will display on the Run Screen after

the initial YSI splash screen. If applicable, install and configure DC power backup

page 45. Verify AC power at the Menus → System →Version menu - page 125.

See Power Fail Menu - page 117 for AC power event log information.

Wire DC Power and backup power

The 5200A may be powered by most 9.0-16.5 VDC source that provides 800mA of

current and is isolated from mains supply by double or reinforced insulation. The

DC power source is user-supplied and can come from a variety of choices including

lead acid, gel cell, or UPS (with 12VDC transformer) external batteries. DC power

wires are terminated at location B on the I/O Board - figure 3.18.

Notes:

The 5200A does not charge batteries. Quality assurance maintenance

-

procedures should be established if batteries are to be fully powered when

used as back-up power source.

Power supply voltage above 16.5 VDC may permanently damage the

-

5200A.

Installation and Wiring

When back up voltage falls below 9.0 volts, the 5200A ceases to operate

-

properly.

Some switching DC power supplies not supplied by YSI may result in

-

noisy readings.

Directions to Wire DC Power/DC Backup Power

WARNING: Disconnect external power to the unit before wiring.

Follow

all manufacturer’s safety and installation instructions when providing power to

the 5200A via a DC power supply.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant

d’effectuer le câblage. Suivez toutes les consignes de sécurité et d’installation du

fabricant lorsque vous fournissez de l’alimentation au modèle 5200A par un bloc

d’alimentation à courant continu.

Perform steps 1 - 3 of 5200A installation.

1.

Open front panel -

2.

page 39.

Feed DC power cable through drilled hole in rubber grommet to location

3.

B. The terminal strip is removable.

CAUTION

: Run high and low voltage cables through separate bulkhead and

conduit.

Terminate DC power wires to pins (-) and (+) according to manufacturer

4.

instructions and any applicable local electrical codes.

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5200A is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.18

Installation and Wiring