Close front panel control board, Step 3 ground 5200a – YSI 5200A User Manual

Page 21

Advertising
background image

YSI 5200A
40

YSI 5200A
41

Close Front Panel

Tighten bulkhead fittings after connecting all 5200A wiring. (Hand tighten

1.

fittings around rubber grommet. Tighten bulkhead fittings into 5200A

enclosure with wrench being careful to not overtighten.)

Follow information in

2.

Sealants, Desiccants and Securing the Monitor

Section - page 61 to safeguard against damage to internal electronic

components.

Make certain the gasket in the front panel lid is in place and not twisted

3.

or damaged.

Replace front panel making sure that ribbon cable is not trapped in the

4.

gasket channel before inserting mounting screws.

Tighten mounting screws, making sure not to cross thread. The screws are

5.

stainless steel, and the receiving threads are brass. Do not over-tighten!

WARNING: The 5200A utilizes sensitive solid-state devices that can be dam-

aged by static shock. Installers must observe accepted ESD (Electro-Static

Discharge) procedures while connecting cabling to the 5200 I/O plate or to

any other internal component or damage may result.

Control Board

The control board is mounted underneath the I/O board. The serial to ethernet

device and lithium battery are installed on the control board. Complete installation

instructions, including accessing the control board, are provided with the optional

serial to ethernet device. Figure 3.15 shows the control board and I/O board removed

from the 5200A enclosure.

Installation and Wiring

Figure 3.15

Step 3 Ground 5200A

WARNING

: Ground the 5200A to avoid possible electrical shock or damage

to the equipment.

AVERTISSEMENT : mettez le modèle 5200A à la terre afin d’éviter tout risque

de choc électrique ou d’endommagement de l’équipement.

Ground the 5200A at location E on I/O Board - figure 3.16. Other 5200A system

components can be grounded at locations F and G. Titanium ground rods can

also be used to ground peripheral equipment.

In addition to grounding the 5200A, all tanks should be electrically grounded via

a ground probe.

Directions to Ground 5200A

WARNING: Disconnect external power to the unit before wiring.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer

un câblage quelconque.