After the dwell time expires, if the value is still in control range, the control relay

will reenergize. A dwell time must be configured in the timed mode or the control

relay will never de-energize regardless of sensor value.

Temperature Compensation

Conductivity values vary with temperature. Conductivity values are normally

compensated to 25°C. An adjustment of approximately 1.91% per °C is common.

Enter “0.00%” at the temperature compensation menu to disable temperature com-

pensation. “*” is displayed next to a conductivity value at the Run Screen to indicate

temperature compensation is enabled.

Directions to Configure Conductivity Sensor

Connect probe assembly -

1.

page 60.

Enter Conductivity Sensor Setup menu

2.

Menus →Sensor Setup→Conductivity

Use

3.

and

to scroll and highlight submenus.

Press

4.

to select.

Configure submenus.

5.

Press

6.

to save configuration and return to Run Screen.

Configuring the 5200A

Notes:

Conductivity set points are always entered in µS (multiply mS value X

-

1000 to obtain µS value).

Only whole numbers can be configured for conductivity set point, control,

-

and alarm values.

See 30 Second Sensor Hold Off

-

- page132.

Active dosing timers are reset after systemwide events. Systemwide events

-

de-energize relays. Example - conductivity dose up relay is configured

for two (2) minutes. Energized dose up relay de-energizes when 5200A

goes into Svr Hold. The dose up relay had been energized for 15 seconds

before Svr Hold occurred. Svr Hold time is one (1) minute. After Svr Hold

time expires, the control system is reset. If sensor is reporting control

low condition the dose on time will be for two (2) minutes and not the

remaining 45 seconds of the dose up time prior to the Svr Hold.

Conductivity sensor log is reported in Siemens (S). 1 S = 1,000

-

millisiemen(mS) = 1,000,000 microsiemen (µS). The conductivity sen-

sor log resolution is 0.00. For example, 68.37 mS is stored as 0.07 in the

sensor log and 5960 µS is stored as 0.00.

Conductivity Control Low Range Low message remains at the

-

menu when the conductivity system is disabled when a control low relay is

energized. Power cycle the 5200A to clear the control low range message

at the Msg’s menu. AquaManager data log displays 5200A conductivity

values in µS to 0.00 resolution.

DO

Menus →Sensor Setup→DO

See Set points, Control, and Alarms - page 128 before configuring the DO system.

Salinity & Temperature Compensation for DO Readings

In order to report accurate DO values, the salinity and temperature of the monitored

water must be known. Temperature is measured by the temperature sensor located

on the probe assembly. Salinity is either calculated by the conductivity sensor (see

Salinity - page 145) or by entering a user defined ppt value in the DO menu. If the

salinity system is enabled, then the DO salinity menu is not available to configure.

Notes:

When conductivity is enabled, salinity is ALWAYS calculated by the con-

-

ductivity sensor and this value is used to calculate DO mg/L.

Disabling conductivity requires that a ppt value be entered in the 5200A

-

DO sensor menu. In this menu, enter the salinity value of the water being

monitored. The acceptable salinity input range is 0-50 ppt.

Configuring the 5200A