Go to Sensor Setup Menus → Sensor Setup Aux 1-2 -

6.

page 133 to enable

and configure set points, control, alarm, relays, and hold off time.
Press

7.

to save configuration.

Note:

The alarm low value must be ≥ the min value. The alarm high value must

-

be ≤ the max value.

Data Logging

Menus → System →Data Logging

The 5200A saves information in the calibration, sensor, and event logs. Press

at the Run Screen to view the 25 most recent records for each log. The 5200A sensor

log can store 1000 records. The calibration log can store 500 records. The event log

can store 1250 records. Once a 5200A log is full, it will begin overwriting the oldest

records with new records (data wrapping).

Notes:

- Use AquaManager to store and analyze 5200A logs. To avoid gaps in data

use polling intervals based on the user defined data logging configuration.

Set the data log interval so that analyzing the data is meaningful.

- AquaManager users can view Configuration events in the

Configuration log. Configuration

events are created when certain 5200A

configuration changes are made. The configuration log is not accessible at

the 5200A. See AquaManager Online Help for additional information.

Data Logging

Menus → System →Data Logging→Data Logging

The 5200A sensor log stores records for sensors and aux systems. Sensor logs are

time and date stamped based on the user configured logging interval (1-255 min-

utes/hours).

Configuring the 5200A

Enter Data Logging menu

1.

Menus → System →Data Logging→Data Logging

Use

2.

and

to scroll and highlight submenus.

Press

3.

to select.

Enable and configure submenus.

4.

Press

5.

to save configuration and return to Run Screen.

Press

to view the 25 most recent sensor logs.

Notes:

Temperature values are stored in °C regardless of display format.

-

Aux digital “closed” is logged as Aux 0.00. Aux digital “open” is logged

-

as Aux 1.00.

The 5200A sensor log records 0.00 resolution for aux analog inputs regard-

-

less of the selected display format.

DO values are stored in mg/L & %Sat regardless of selected display

-

format.

Sensor logs record the default system name and not the user configured

-

system name.

Changes to the data logging interval take effect after the expiration of the

-

interval being changed. Example - if the interval is changed from one

hour to one minute, one minute logging will begin after the expiration of

the current one hour interval. Power cycle the 5200A to make a change

to the data logging interval occur immediately.

Menu configurations are saved to the system by pressing

-

to

return to the Run Screen and not when the change is made in the menu.

Exception - Date and time and clear data logs are updated immediately
to the system, not when

is pressed.

The sensor log contains values for all sensor and aux systems regardless

-

if the system is disabled.

5200A sensor log values for disabled systems and systems without sen-

-

sors are not valid.

The Conductivity sensor log is reported in Siemens (S). 1 S = 1,000

-

millisiemen(mS) = 1,000,000 microsimen (µS). The conductivity sensor

log resolution is 0.00. For example, 68.37 mS is stored as 0.07 S in the

sensor log and 5960 µS is stored as 0.00 S.

AquaManager data log displays 5200A conductivity

-

values in µS to 0.00

resolution.