Non-Networked (Stand Alone) 5200As

Non-networked (stand alone) 5200As must have the termination relay enabled. In

addition, non-networked (stand alone) 5200As must be configured as a master in

the System Communication menu with no slave nodes enabled.

Configure AUX

Menus → System → Configure Aux 1-2

The 5200A has two auxiliary (Aux) inputs. Aux inputs can be configured as digital or

analog (0-1VDC, 0-5VDC, or 4-20 mA) inputs. Configure suffix, minimum (Min),

maximum (Max), and display format menus for the analog inputs. Go to Sensor

Setup Menus → Sensor Setup Aux 1-2 - page 133 to enable Aux channel and configure

control/alarm settings. Both digital and analog directions are on the following pages.

Submenu functionality is provided after the configuration directions.

Probe Name

Menus → System → Configure Aux →Aux 1-2→Probe Name

Use the alpha/numeric keypad to name aux inputs. Configured name(s) appear

throughout menus and at Run Screen - figure 4.6, page 101.

Note:

Aux names can also be configured at Menus → System → Name Devices/

-

Sensors.

Input Type

Menus → System → Configure Aux →Aux 1-2→Input Type

Select the input type for the device(s) wired to aux system(s). Aux Sensor Setup

(page 133) menu selections are based on the aux input type configured at the Sys-

tem menu.

Configuring the 5200A

Digital

Menus → System → Configure Aux →Aux 1-2→Input Type→Digital

Use the digital input when wiring an external device to control or alarm when the

contacts of the external switch are open or close. A common application would be

to detect high or low water levels using a float switch.

Notes:

Suffix, Min, Max, and Display Format are not available for digital in-

-

puts.

In the 5200A sensor log, aux digital “closed” is logged as Aux 0.00. Aux

-

digital “open” is logged as Aux 1.00.

Wire aux digital devices -

1.

page 49.

Enter Aux menu

2.

Menus → System → Configure Aux →Aux 1-2

Use

3.

and

to scroll and highlight submenus.

Press

4.

to select.

Configure digital input type and name.

5.

Press

6.

to save configuration and return to Run Screen.

Go to Sensor Setup Menus → Sensor Setup Aux 1-2 -

7.

page 133 to enable

and configure control/alarm on condition, control/alarm relay, and hold

off time.
Press

8.

to save configuration.

Configuring the 5200A