Sensor troubleshooting, Appendix 1 - unpacking and inspection, Appendix 2 - troubleshooting – YSI 5200A User Manual

Page 95

Advertising
background image

YSI 5200A
188

YSI 5200A
189

Appendix 1 - Unpacking and Inspection

Inspect the outside of the shipping carton for damage. If damage is detected, contact

the carrier immediately. Remove the instrument from the shipping container. Be

careful not to discard any parts or supplies. Confirm that all items on the packing

list are present. Inspect all assemblies and components for damage. Save the original

packing carton. Carriers typically require proof of damage due to mishandling.

The following components are included with the purchase of the 5200A instru-

ment:

5200A instrument

006515 Flange Mounting Kit

006506 Desiccant Kit

605876 Instruction Manual

605230 Compression Plugs, 2 each

605229 Compression Plugs, 2 each

605804 Compression Plugs, 1 each

655384 Dual Male RJ-45 Cable Assembly

655383 Adapter, DB-9 to RJ-45

605880 AC Power Cord, 115 VAC U.S. Plug (AC version only)

605944UL Pluggable Socket Connector for AC power, 3 pole (AC Version only)

506939UL Pluggable Socket Connector for 12 VDC Power, 2 pole

605941UL Pluggable Socket Connector for RS-485, 5 pole

605943UL Pluggable Socket Connector for relay, 12 pole

605940UL Pluggable Socket Connector for Auxiliary inputs, 3 pole

If any parts are damaged or missing, contact YSI or your YSI representative im-

mediately.

Unpacking and Inspection

Appendix 2 - Troubleshooting

Sensor Troubleshooting

When troubleshooting sensors, disable the control/alarm relays.

-

Verify 5200A wiring and configuration.

-

Issue

possible cause

possible solution

inaccurate or fluctuating

sensor values

improper placement of probe

assembly

See Location Considerations

- Step 2 Locate and Install

Probe Assembly - page 33.

improper placement of

5200A

See Location consider-

ations - Step 1 Install 5200A

- page 27.

bad sensor

bad DO membrane

recalibrate, change mem-

brane, clean sensor, test

sensor outside system

stream in solution of

known value, replace sensor

sensor out of calibration

See Calibration - page 72.

DO elevation configuration

incorrect

Correct DO elevation at

Menus → System → Sensor

Setup→DO→Elevation

menu

Conductivity, DO, and pH

values are effected by temper-

ature. If conductivity, DO,

and pH values are incorrect,

verify that the temperature

probe is working correctly.

If the temperature probe is

reporting incorrect values,

conductivity, DO mg/L, and

pH values are not reliable.

See temperature compen-

sation and conductivity -

page138.

See DO and temperature -

page139.

“over/under” sensor

readings

bad sensor

bad membrane

test sensor outside system

stream in solution of

known value, clean sensor,

recalibrate, replace sensor

and/or membrane if neces-

sary

Troubleshooting