Removing the AUX Power Connector
Tools Required
•
Socket, 9100-3060-AF
•
Large diagonal cutters
Setup
•
This procedure should be conducted on a static mat.
•
Wear a wrist strap to ground during the procedure.
•
Complete these procedures in the following order before performing this procedure:
1.
“Removing the Handle” on page 63
2.
“Removing the Printer Module” on page 64
3.
“Disassembling the Main Unit” on page 66
4.
“Removing the Defibrillator Boards and Capacitor Bank Modules” on page 88
AUX Power Connector
1
Disconnect AUX Power
Connector cable from
harness.
2
Using socket, 9100-3060-AF,
remove lock nut that is
securing the connector to the
housing.
Note: During reassembly,
apply a drop of loctite before
secure the lock nut.