9650-001356-01 Rev. D
ZOLL X Series Service Manual
131
Appendix B
Overview
This appendix includes simulators and settings that may be used to assess the performance of the NIBP module.
Readings should be obtained within +/- 10 mmHg of the expected measurement result. The following simula-
tors are used in this chapter:
•
Fluke Biomedical BP Pump/Bio-Tek Instruments BP Pump
•
Fluke Biomedical BP Pump/Bio-Tek Instruments BP Pump2
•
Fluke Biomedical BP Pump/Dynatech Nevada Inc. CuffLink
•
Clinical Dynamics Smart Arm
NIBP Simulator Performance
FastBP
A dual lumen hose is required for FastBP.When you use FastBP, the reading is obtained during cuff inflation.
StepBP
StepBP is compatible with both single and dual lumen hoses.When you use StepBP, the reading is obtained
during step deflation.