0 temperature test, Temperature test – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 50
Advertising
C
HAPTER
1
M
AINTENANCE
T
ESTS
44
9650-001356-01 Rev. D
25.0 Temperature Test
Equipment
IBP Temperature Simulator (or equivalent)
Do this...
Observe This...
Pass/Fail
25.1 Connect the temperature simulator to the
X Series T1 channel.
25.2 On the IBP Temp Simulator, Select 98.6
Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.
The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.
o
o
25.3 Repeat steps 1 and 2 of this procedure for the
T2 channel.
25.4 Disconnect cable.