C

HAPTER

1

M

AINTENANCE

T

ESTS

44

9650-001356-01 Rev. D

25.0 Temperature Test

Equipment

IBP Temperature Simulator (or equivalent)

Do this...

Observe This...

Pass/Fail

25.1 Connect the temperature simulator to the

X Series T1 channel.

25.2 On the IBP Temp Simulator, Select 98.6

Degree F (36 Degree C) or a setting as close
as possible as your simulator allows.

The displayed temperature value is
within 0.2 F or 0.1 C of the
simulator setting.

o

o

25.3 Repeat steps 1 and 2 of this procedure for the

T2 channel.

25.4 Disconnect cable.