Spo2, spco, and spmet test (for spo, 0 spo2, spco, and spmet test (for spo, Feature with spco, or spco and spmet) – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 37: 0 spo, Spco, and spmet test (for spo

Advertising
background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

31

15.0 SpO

2

, SpCO, and SpMet Test

(for SpO

2

feature with SpCO, or SpCO and SpMet)

Note:

Use this test if your unit has SpO

2

and SpCO, or SpO

2

, SpCO and SpMet. If you do not

have the SpMet feature on the unit, disregard the SpMet values in the test below.

Equipment

Fluke Index 2 SpO

2

Simulator (or equivalent), Masimo Rainbow 3 ft BioTek Test Sensor

(Masimo P/N 3379) or Masimo Rainbow 3 ft Direct Connect Optical Test Sensor

Test Setup

Disconnect any ECG cable or IBP probes from the X Series unit.

Do This...

Observe This...

Pass/Fail/N/A

15.1 Power on the X Series unit without the

SpO

2

sensor attached.

Verify that the

SpO

2

numeric

displays INITIALIZING, then
goes blank.

15.2 Connect the

SpO

2

sensor to the X

Series unit.

Verify that

SpO

2

numeric

displays SEARCHING.

15.3 Set the simulator to 75%

SpO

2

and 50

BPM. Attach the

SpO

2

sensor to the

simulator.

Verify that after a brief period of
initialization,
The

SpO

2

numeric displays an

SpO

2

value between 72.1% and

78%.
The

SpCO

numeric displays an

SpCO

value between 13.5 and

21.5.
The

SpMet

numeric displays an

SpMet

value between 5.5 and

9.5.

Verify that the displayed pulse
rate reads between 44.5 and 55.5
BPM.

o

o

o

o

o

o

o

o

o

15.4 Set the simulator to 75%

SpO

2

and 230

BPM.

Verify that after a brief period of
initialization,
The

SpO

2

numeric displays an

SpO

2

value between 72.1% and

78%.
The

SpCO

numeric displays an

SpCO

value between 12.8 and

20.8.
The

SpMet

numeric displays an

SpMet

value between 5.2 and

9.2.

Verify that the displayed pulse
rate reads between 222.7 and 236
BPM.

o

o

o

o

o

o

o

o

o