ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 98

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

92

9650-001356-01 Rev. D

11

Cut the tie wrap securing the
Power Monitor to Pace cable
(black and red coupled wire)
and disconnect the cable (via
the red connector) in
illustrated area.

12

Remove four (4) screws from
the frame support in the
areas illustrated.

Note: During reassembly,
torque screws to 6 in-lbs.

13

Lift up on the ECG assembly
and remove red and white
MFE cable (via the white
connector) from the ECG
board. Remove ECG
assembly.