19.0 IBP Test
Tools Needed
IBP Simulator (with accuracy better than +/- 1 mmHg), or equivalent
Setup
Connect the test cables to IBP patient simulator.
Do This...
Observe This...
Pass/Fail
19.1 Connect the IBP patient simulator to the X
Series P1 channel.
19.2 Set the IBP patient simulator to 0 mmHg.
19.3 On the X Series unit, press > IBP.
19.4 On the X Series unit, Press > P1->0<-.
Display of 0/0(+/-2) in the P1 box
at the bottom of the display (P1 is
zeroed).
The waveform for channel 1 is
present.
o
o
o
o
19.5 Exit the menu to see the waveform.
19.6 Set the IBP patient simulator to 200
mmHg (or as close to 200 mmHg as your
simulator allows).
Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).
o
o
19.7 Repeat steps 1 through 5 of this procedure
for P2 and P3 channels.
19.8 Disconnect IBP cable.