0 ibp test, Ibp test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 42

Advertising
background image

C

HAPTER

1

M

AINTENANCE

T

ESTS

36

9650-001356-01 Rev. D

19.0 IBP Test

Tools Needed

IBP Simulator (with accuracy better than +/- 1 mmHg), or equivalent

Setup

Connect the test cables to IBP patient simulator.

Do This...

Observe This...

Pass/Fail

19.1 Connect the IBP patient simulator to the X

Series P1 channel.

19.2 Set the IBP patient simulator to 0 mmHg.

19.3 On the X Series unit, press > IBP.

19.4 On the X Series unit, Press > P1->0<-.

Display of 0/0(+/-2) in the P1 box
at the bottom of the display (P1 is
zeroed).

The waveform for channel 1 is
present.

o

o

o

o

19.5 Exit the menu to see the waveform.

19.6 Set the IBP patient simulator to 200

mmHg (or as close to 200 mmHg as your
simulator allows).

Display of 200/200(+/-5) in the P1
box at the bottom of the display as
xxx/yyy where yyy=diastolic (P1
diastolic pressure is 200 (+/-5)).

o

o

19.7 Repeat steps 1 through 5 of this procedure

for P2 and P3 channels.

19.8 Disconnect IBP cable.

Advertising