ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 96

Advertising
background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

90

9650-001356-01 Rev. D

WARNING! This unit may contain lethal voltages. You MUST completely discharge the high voltage capacitor

before removing from unit. DO NOT SHORT THE TRERMINAL ENDS OF THE CAPACITOR.

6

Discharge the capacitor by
bleeding out the excess
voltage using a resistor with
values of approximately 5k
Ohms, (25 watts for 10-20
seconds). Counting from the
right connect, apply the
resistor across the second
and forth via as shown in
picture. Measure the voltage
of the capacitor on the same
two vias to verify that the
capacitor is discharged.

7

Remove the four (4) screws
from the processor board in
the areas illustrated.

Note: During reassembly,
torque screws to 6 in-lbs.

Advertising