Removing the Defibrillator Boards and Capacitor Bank Modules
9650-001356-01 Rev. D
X Series Service Manual
93
14
Remove the 5 screws
securing the ECG board to
bracket, and then separate
the board from the bracket.
Note: During reassembly,
torque screws to 6 in-lbs.
15
Remove the Bridge/Pace to
ECG cable from the ECG
board in the two areas
illustrated.
Note: During reassembly,
match the designator on the
flex cable and the ECG
board.
16
Remove two (2) screws
securing the capacitor to
housing, then lift capacitor
out.