Removing the Defibrillator Boards and Capacitor Bank Modules

9650-001356-01 Rev. D

X Series Service Manual

93

14

Remove the 5 screws
securing the ECG board to
bracket, and then separate
the board from the bracket.

Note: During reassembly,
torque screws to 6 in-lbs.

15

Remove the Bridge/Pace to
ECG cable from the ECG
board in the two areas
illustrated.

Note: During reassembly,
match the designator on the
flex cable and the ECG
board.

16

Remove two (2) screws
securing the capacitor to
housing, then lift capacitor
out.

Note: During reassembly,
torque screws to 6 in-lbs.

Advertising