ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 93
Advertising
Removing the CP Carrier Module
9650-001356-01 Rev. D
X Series Service Manual
87
18
Disconnect the top flex cable
connectors from SPO
2
module, then remove cable
and ferrite.
19
Remove the screw from the
flex cable, then disconnect
flex cable from the
connectors in two locations.
Note: During reassembly,
torque screws to 4 in-lbs.
20
Remove 4 screws from the
SpO
2
module and remove it.
Note: During reassembly,
torque screws to 4 in-lbs.
Advertising