ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 93

Advertising
background image

Removing the CP Carrier Module

9650-001356-01 Rev. D

X Series Service Manual

87

18

Disconnect the top flex cable
connectors from SPO

2

module, then remove cable
and ferrite.

19

Remove the screw from the
flex cable, then disconnect
flex cable from the
connectors in two locations.

Note: During reassembly,
torque screws to 4 in-lbs.

20

Remove 4 screws from the
SpO

2

module and remove it.

Note: During reassembly,
torque screws to 4 in-lbs.

Advertising